CyRISBio is launching a multi-brand qualification service for your pipetting robots

With a dozen liquid handling robots from various brands (Analytik Jena, Tecan, Hamilton, Beckman, etc.), maintenance to schedule, and tests that cannot wait, verifying the performance of each robot is an essential task that can quickly become complex and time-consuming.

Who performs the checks? How can you maintain consistent tracking? How can you limit the time required from laboratory engineers and scientists?

Maintaining a robot is not the same as qualifying it

Even when well-maintained, a robot can experience dispensing drift without any visible signs. If a discrepancy in a few channels is only discovered after a screening, you must reanalyze samples, use more reagents, and remobilize equipment and teams. A small deviation is enough to call an entire series of results into question.

That is the whole difference between maintenance and qualification. Maintenance ensures the robot is serviced. Qualification answers a different question: is the robot dispensing correctly, according to your protocol and acceptance criteria?

It all starts with how you work: what volumes are you dispensing, and with which solutions? A robot is never qualified in the abstract, but rather for the lab's specific application. This tailored approach, inherent to custom automation, is what guides everything we do.

A service for your entire fleet

To meet this need, CyRISBio is opening its qualification service to all laboratories, even for systems we didn't supply or install. This service draws on expertise our technical team already uses every day, performing comparable checks on the pipetting robots they maintain.

Frédéric Vachot, Technical Manager at CyRISBio, and his team master the software for each device. Their extensive experience with liquid handling systems allows them to adapt to diverse configurations, particularly multichannel heads and independent channels.

For a heterogeneous fleet, the advantage is having a single partner for all verifications and their tracking. Suppliers remain responsible for the maintenance of their robots; we verify that each one meets your criteria.

Our three-step intervention

Our intervention follows three steps:

1. Volumetric verification and leak testing

We check the accuracy of volumes, aspiration, and dispensing, as well as well-to-well consistency. Volumetric verification is based on colored reagents dispensed into microplates, the absorbance of which is then measured. The results are compared against laboratory criteria. Depending on the chosen indicators, these may include mean volume, relative bias (accuracy), coefficient of variation (CV, precision), and Z-score distribution.

A leak test completes the check: it looks for any leaks that could affect aspiration or dispensing.

These checks are most effective when they frame maintenance. Before the supplier's intervention, they establish a documented baseline. After the intervention, they verify that the measured performance once again meets the agreed-upon criteria.

2. Correction at the source

When a discrepancy appears, we trace it back to its origin: the head, a channel, or a seal. The correction focuses on the faulty element, and the robot is returned to a compliant state. The points of contact and the procedure to follow in the event of a discrepancy are defined during the preparation phase.

3. Long-term tracking

You receive the results of the checks along with a qualification report. Their transmission and recording, particularly in your tracking software, are organized with your teams according to your tools and procedures.

Performed annually using the same protocol, these checks build a comprehensive qualification history. If your lab uses control charts, repeated results can be plotted against defined thresholds to help spot gradual drift or recurring deviations. These trends can also guide investigations and optimize maintenance scheduling. In this way, the checks complement preventive maintenance while supporting your quality process and full traceability.

A simple framework agreement

Outsourcing these checks should free up time for research technicians and engineers while limiting administrative burden. The service is therefore organized around an annual framework agreement, which defines the scope of interventions and billing terms.

·      Streamlined requests: Once the framework is in place, you can schedule planned interventions without restarting the entire procurement process every time. Monthly billing can also be arranged based on your preferred terms.

·      Predictable budgeting: Knowing the equipment covered, the planned number of qualifications, and the associated rates helps you set your annual budget with confidence. Annual volume and advance planning also make it possible to consolidate site visits and keep service costs under control. Provisions for additional or urgent interventions are included directly in the contract.

·      Scheduling tailored to your needs: Checks integrate seamlessly with preventive maintenance, but they can also be performed on demand—before a major campaign, following a repair, or when recommissioning a robot after an extended shutdown. We recommend at least an annual check, with the actual frequency determined alongside you based on your specific usage, maintenance schedules, and quality requirements.

Each month, a review with our technical manager takes stock of progress: which services were completed, which results require attention, and what still needs to be scheduled. You benefit from a dedicated point of contact who knows your equipment inside and out, rather than starting from scratch with every request.

This solution is particularly well-suited for fleets requiring recurring checks. It may also be justified for a single robot if the frequency of checks demands it.

Is your next maintenance or campaign coming up?

Let us know the number and brands of your pipetting robots, as well as your next deadline. Together, we will define the protocols, the inspection schedule, and the service terms.

A single partner for your verifications, fully documented results, and ongoing support tailored to your equipment. Let's talk about your pipetting robots!

‍

Have an automation project in mind?

Describe your process to us, and we will propose a custom solution.

Contact us

Foire aux Questions

Voit notre FAQ

What is the average lead time from initial contact to the delivery of a final solution?

The average lead time for small-scale automation is around 3 months. For more complex laboratory automation, it varies between 4 and 10 months, with an average of 6 months.

Does CyRISBio partner with specialized manufacturers for its pipetting solutions?

Yes, CyRISBio distributes, among other things, Analytik Jena's CyBio automated pipetting platforms, which are used for laboratory liquid handling.

What does CyRISBio's sample management offering include?

Our sample management offering covers automated low-temperature storage (biobanking), incubators, and storage hotels, as well as associated consumables such as compatible microtubes and racks, notably through our partner LVL-Technologies.

What does CyRISBio's laboratory automation offering consist of?

CyRISBio provides automation solutions for liquid handling and dispensing, as well as the integration of robotic platforms to automate complete laboratory workflows, notably leveraging Analytik Jena's CyBio liquid handling platforms.

Does CyRISBio manufacture its own equipment or distribute partner brands?

CyRISBio designs custom solutions and also distributes equipment from recognized specialized manufacturers in laboratory automation, in order to offer its clients the solution best suited to their process constraints.

Which industry sectors does CyRISBio support?

CyRISBio supports laboratories across several sectors: pharmaceuticals, cosmetics, chemicals, food processing, and biology, with automation solutions tailored to each type of process.

Who is CyRISBio?

CyRISBio is a French company specializing in laboratory automation, founded in 1999. Based in Chaponost, in the Lyon region, it designs and distributes robotic and instrumented solutions for laboratory pipetting and sample management.

We want to automate a manual process but have technical concerns regarding our workflows; can we conduct a POC with you?

Yes. Transitioning from a manual laboratory process to an automated one can indeed raise a number of questions and challenges. We will work on providing automation sub-components to address these issues. These will be handled through micro-developments that will strengthen the final solution.

Is there a cost for being put in touch with your project engineers to evaluate a solution for our automation needs?

No! The initial introductory meeting often takes place via video conference. This allows us to quickly and easily introduce ourselves and assess our mutual interest in moving forward. If both parties decide to work together, the project engineer begins their work, and billing is triggered upon the successful completion of the operation.

Do you need detailed specifications to start an automation project?

The more detailed your laboratory automation requirements are, the less room it leaves for our creativity. Ideally, you should define your needs and outline your processes clearly; the final specifications will be established after a few exchanges with the project engineer in charge.

‍

Can you mix tube formats in an automated sample storage Biobank?

Yes, absolutely. It is possible to mix different types of cassettes with varying heights within the automated sample storage system. This allows you to combine everything from tiny 300 µL tubes to 2 mL tubes.

What is the maximum capacity of your -80°C Biobank?

LiCONiC builds custom Biobanks tailored to your required capacity. There are no capacity limits. Examples of previously manufactured systems include the STC 101k0-ULT.

This automated -80°C sample storage system has a capacity of 3.6 million 1 mL tubes, or 38,400 racks of 96 tubes. By using high-density (HD) racks, the capacity increases by over 40% to reach 5.5 million tubes.

What is the smallest capacity of your -80°C biobank?

The smallest automated -80°C sample storage system, the STT 1.5K ULT, has a capacity of 49,920 1 mL tubes, or 520 96-tube racks. By using smaller 300 µL tubes, for example, the capacity increases to 122,800 tubes.

Where are LVL-Technologies 2D microtubes manufactured? Is the supply chain secure and reliable?

The Safe® range of 2D microtubes is designed, manufactured, and tested in Europe within controlled industrial environments that meet the rigorous demands of scientific and biotechnological applications.
This European production ensures consistent quality, full batch traceability, and strict compliance with international standards.

Control over the production and supply chain allows LVL-Technologies to ensure reliable and long-term product availability, even for large-scale projects or long-term biobanking initiatives.
This approach protects users against supply chain disruptions and ensures the continuity of critical research and sample storage operations.

How can I increase storage space in my -80°C freezers?

Significant space savings can be achieved by replacing internal thread tubes with external thread 2D microtubes.
Thanks to the cap design, where the thread sits on the outer top edge of the tube, this configuration allows for reduced height and therefore increased storage density.
Furthermore, switching from 9×9 or 10×10 cryoboxes to an SBS format saves additional space while paving the way for future biobank automation.
Finally, integrating the Hi-Thèque software helps regularly optimize stock organization while ensuring complete and secure sample traceability, both on-site and remotely.
Our teams are at your disposal to assist you with all initiatives aimed at sustainably optimizing your storage capacity.

Are the 2D microtubes in the Safe range guaranteed RNase/DNase-free?

Yes. The 2D microtubes in the Safe® range are guaranteed RNase-free and DNase-free.
They are manufactured under controlled, sterile conditions and undergo strict quality procedures, ensuring their compatibility with molecular biology and genomics applications, as well as the storage of high-value scientific samples.

How can I effectively manage the storage of my 2D microtubes in -20°C or -80°C freezers or liquid nitrogen tanks?

The Safe® range of 2D microtubes is designed to be stored in standard SBS racks or cryoboxes.
LVL-Technologies also offers customized metal storage boxes, compatible with the main models of refrigerators, freezers, and liquid nitrogen tanks available on the market.
Individual identification via 2D DataMatrix code ensures fast, reliable, and secure sample management, facilitating inventory and storage space optimization.
For advanced management, the use of automated scanning systems and tracking software is highly recommended.

In this regard, the Safe® range from LVL-Technologies includes a new generation of scanners capable of reading all types of 2D microtubes, either individually or in SBS racks and/or 9×9 or 10×10 cryoboxes.

Is it possible to evaluate the chemical inertness of the 2D microtube plastic with respect to biological substances?

Yes. The 2D microtubes from the LVL-Technologies Safe® range are manufactured from medical-grade polypropylene (USP Class VI), selected for its high chemical inertness, very low extractables, and compatibility with demanding biological applications.
These materials are perfectly suited for the long-term storage of sensitive biological samples and limit interactions with DNA, RNA, proteins, or other analytes, even under ultra-low temperature conditions over long periods.
Quality and compliance evaluation tests have been conducted to meet the standards of research laboratories and biobanks. Detailed results are available in the LVL-Technologies Technote, upon request.

Should you choose external or internal threading for ultra-low temperature storage?

For ultra-low temperature storage (-80°C and below), LVL-Technologies recommends using external thread 2D microtubes.
‍

This configuration significantly reduces the contact surface between the sample and the inside of the cap, thereby limiting the risk of cross-contamination, material transfer, or unwanted physical-chemical interactions.
These risks are particularly high with internal thread tubes during agitation or centrifugation, where liquid samples can rise and pool along the internal thread.
A comparative study detailing these phenomena is available in the LVL-Technologies Technote, accessible to all our users upon request, in either paper or electronic format.